Why is collimation not holding?
If you’ve followed the HyperStar collimation steps carefully — established the correct mechanical starting position, performed rough collimation, and refined it using in-focus stars — the system should hold alignment. HyperStar collimation is not something that needs to be repeated every session.
When customers contact us because collimation seems to drift or never quite hold, the cause is often mechanical rather than optical. Before repeatedly adjusting the collimation screws, it’s worth checking that the HyperStar and its mounting components are secure and that the telescope itself is properly aligned.
1. The HyperStar Was Not at the Right Starting Position
Collimation adjustments work predictably only when the HyperStar begins from the correct mechanical baseline. The top and bottom mounting plates should start flat against each other, with no residual tilt from previous adjustments. If the plates are uneven or the screws have been tightened inconsistently, collimation may appear to improve temporarily without remaining stable.
For example, one screw may have been tightened significantly before the others, or an apparent gap between the plates may actually be uneven tilt rather than the normal bevel clearance. These conditions can leave the adjustment screws with uneven tension, making it harder to achieve stable alignment.
If you suspect this is happening, return the HyperStar to its correct starting position before attempting further adjustments.
See Step 1: HyperStar Collimation Starting Position for detailed instructions.
2. The Collimation Screws Are Not Tightened Enough
HyperStar collimation uses opposing push–pull screw pairs to adjust the tilt of the lens assembly. The short screws pull the mounting plates together, while the taller screws push against the lower plate to raise one side. Both screws in each pair must work together to maintain the adjustment.
If a pull screw is loosened without adequately tightening its corresponding push screw, or several screws are loosened without maintaining tension elsewhere, the assembly can settle or shift. Skipping the final tightening after fine collimation can also allow the alignment to change during transport or as the equipment is handled.
Make small adjustments, maintaining tension between the opposing screws. Once collimation is complete, snug the screws with a screwdriver to help them hold their position. Properly tightening the collimation screws will not damage the HyperStar and helps preserve alignment between sessions.
3. The HyperStar Is Overtightened on the Telescope
The HyperStar should be threaded onto the telescope until the threads naturally bottom out. It does not need an extra tightening twist to remain secure.
Overtightening can cause the threads to bind, particularly in cold conditions when metal components contract. The HyperStar may become difficult to remove, and the secondary assembly can rotate during removal, potentially disturbing the telescope's alignment.
If the HyperStar is difficult to remove or collimation changes after reinstalling it, check whether it has been overtightened to the telescope or whether the secondary assembly has shifted. When removing and reinstalling the HyperStar, handle the components carefully to avoid rotating the secondary mirror assembly.
4. The Rotation Screws or Other Components Are Loose
The three HyperStar rotation screws, identifiable by their large white bushings, hold the HyperStar against its mounting rings while allowing the lens and camera to rotate together for framing. If these screws are loose, the HyperStar can sag under gravity, causing the camera and lens assembly to tilt and lose collimation.
Check that all three rotation screws are secure. When rotating the HyperStar, loosen the screws only enough to allow movement. After framing your target, tighten the topmost screw first to pull the HyperStar up against gravity and seat it flat against the mounting rings, then tighten the other two screws.
Other components around the telescope's corrector plate can also affect alignment. A loose inside baffle ring, outer retaining ring, or slipping gasket can allow the secondary assembly to rotate or shift slightly during HyperStar removal and reinstallation. This can make collimation appear inconsistent between sessions, even if the HyperStar's own collimation screws have not moved. In some cases, worn or slippery gaskets may need to be replaced.
5. The Corrector Plate Is Not Centered
Corrector plate centering is an important part of HyperStar alignment that is easy to overlook. The HyperStar must be optically centered over the telescope's primary mirror, which means the corrector plate — or, more precisely, the secondary holder mounted in it — must be optically centered as well.
Mechanical centering does not necessarily guarantee optical centering. If the corrector plate is off-center, collimation adjustments may behave unpredictably. Stars may improve on one side of the field while becoming worse on the other, or the adjustments may never seem to produce an even result across the image.
You can check the centering by looking through the front of the telescope with the HyperStar or secondary mirror removed. The secondary holder, its reflection, and the primary baffle should appear as three concentric circles when viewed from the appropriate distance. If they do not, the corrector plate may need to be repositioned.
For more information, see Corrector Plate Centering.
6. The Issue Is Not Actually Collimation at All
Not every star-shape or image-quality problem is caused by misalignment. Incorrect backfocus or spacing, inaccurate focus, calibration problems such as flat-field errors or reflections, and optical elements installed backward can all produce symptoms that might initially be mistaken for poor collimation.
This is particularly worth checking if you purchased a used HyperStar, since a previous owner may have changed the orientation of an optical component. If you suspect that something has been installed incorrectly, contact us at info@starizona.com so we can help verify the configuration.
Focus drift during an imaging session can also make stars appear worse over time, even when the optical alignment has not changed. Similarly, camera sensor tilt can affect star shapes across the field, although it is worth eliminating other possible causes first.
If collimation adjustments do not produce meaningful, repeatable changes in the image, alignment is unlikely to be the root cause. Before making further adjustments, check your focus, backfocus, optical configuration, and calibration.
When to Reach Out
If you are experiencing imaging issues with your HyperStar and are unsure whether collimation is the problem, we're happy to help.
Email us at info@starizona.com with:
A photo of your optical train from the side, showing the camera, adapters, and any filter accessories.
A single raw, unprocessed image showing the star shapes.
These two images usually help us determine whether the problem is collimation, mechanical alignment, spacing, or something else entirely.